Diabetes Autoimmune Test Group
Insulin Antibodies, Islet-cell Antibodies
Test Codes
Mayo: DBS1, Beaker: XDATG, EPIC: LAB6003
Department
Send Outs
Specimen Collection Criteria
Collect (preferred specimen): Two Gold-top SST tubes.
Also acceptable: Two plain Red-top tubes.
Physician Office/Draw Specimen Preparation
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells within one hour of collection. Transfer serum to a plastic transport tube and refrigerate (2-8°C or 36-46°F).
Preparation for Courier Transport
Transport: 4.0 mL serum, refrigerated (2-8°C or 36-46°F). (Minimum: 2.0 mL)
Rejection Criteria
Hemolyzed, icteric, or lipemic samples.
Specimens not collected and processed as indicated.
In-Lab Processing
Let specimen clot 30-60 minutes then centrifuge to separate serum from cells within one hour of collection. Transfer serum to a plastic transport tube and refrigerate (2-8°C or 36-46°F).
Transport: 4.0 mL serum, refrigerated (2-8°C or 36-46°F). (Minimum: 2.0 mL)
Storage
Specimen Stability for Testing:
Room Temperature (20-26°C or 68-78.8°F): 72 hours
Refrigerated (2-8°C or 36-46°F): 28 days
Frozen (-20°C/-4°F or below): 28 days
Specimen Storage in Department Prior to Disposal:
Specimen retention time is determined by the policy of the reference laboratory. Contact the Send Outs Laboratory with any questions.
Laboratory
Sent to Mayo Clinic Laboratories in Rochester, MN.
Performed
Varies.
Results available in 7-11 days.
Reference Range
By report.
Test Methodology
ELISA and Radioimmunoassay.
Clinical Utility
Antibodies to insulin and pancreatic proteins may be used to differentiate type 1 and type 2 diabetes. Radioimmunoprecipitation assays with recombinant antigens assure assay sensitivity and specificity. Testing for multiple markers increases diagnostic sensitivity to 98% and increases positive predictive value of the tests.
CPT Codes
86337, 86341 x3.
Contacts
Send Outs Laboratory – RO
248-551-9045
Name: Send Outs Laboratory – RO
Location:
Phone: 248-551-9045
Last Updated
4/10/2021